70PRO Leak Tester

- General
- Operation
- Technology
The 70PRO Leak Tester
HeMaTech Prüftechnik is known for its robust, durable, and highly accurate leak testers. Users around the world rely on the reliability and service provided by HeMaTech Prüftechnik. Leak testers, often referred to as leak test equipment or simply “leak testers,” are used for leak testing in industrial applications. Whether in the laboratory or in the production of mass-produced parts, our testers can be used in a wide variety of applications. Customers from a wide range of industries use our leak testers, including the automotive industry, medical technology, and electronics manufacturing.
The Advantages of the Leak Tester
The 70PRO is the right choice when you need a rugged and durable leak tester. It stands out for its high measurement accuracy and virtually unbeatable reliability.
The leak tester can store 10,000 data records of test results for later use. Additionally, the data can be exported to a USB flash drive, for example, after each measurement.
The 70PRO can be customized with a wide range of options to suit your specific testing needs.
The Functionality of the Tester
The differential pressure leak testing method used offers many advantages for the user.
The 70PRO's features at a glance:
- Compressed air as a test medium
- Air can quickly seep through leaks
- high measurement accuracy
- reproducible measurements
- Unambiguous measurement results
- Contamination-free testing
- Short cycle times combined with high precision
- Space for 99 test programs
- Storage for 10,000 result records
- Direct export of results to a USB flash drive
- Automated testing sequence
- Can be integrated into plant designs
- Profibus, Profinet, and RS-232 interfaces available
- Display of the measured value and test result
- Variable test pressures and test times for optimal adaptation to the manufacturing of diverse components
- Easy to use thanks to intuitive user guidance
- Service-friendly modular design
Differential pressure measurement technology requires pressure-tight devices and/or sealing elements. HeMaTech can supply you with complete test stations or systems.
The Leak Calibrator for the Leak Tester
Our product line also includes test leaks and a leak calibrator as supplementary accessories. You can use the test leak to quickly and easily check that your leak tester is functioning properly. The leak calibrator is a portable handheld device that allows you to perform reliable calibrations at regular intervals. By performing these checks at any time, you can ensure consistent quality in your measurements.
Calibration
A leak tester is always subject to a certain amount of wear and tear during operation, and electronic components can change their behavior over time. Therefore, a leak tester should be calibrated and maintained regularly.
We recommend having the equipment calibrated every 12 months. This can be done at our workshop or at your location.
HeMaTech Prüftechnik's calibration laboratory is accredited, which means we can also perform DAkkS calibrations.
How the Leak Tester Works
The method is based on measuring the change in pressure within a sealed volume (test specimen) over a specified measurement period, known as differential pressure measurement. A differential pressure transducer with automatic zero-point correction serves as the sensor.
Air escapes through a leak in a sealed volume (test specimen). This causes a pressure drop. In contrast, a reference chamber remains at the original initial pressure. The difference between these two chambers can be measured using a differential pressure transmitter. The pressure difference is a measure of the existing leak and is independent of the test pressure or the prevailing ambient pressure. Inexpensive compressed air is used as the test medium. Due to its low viscosity, the air can quickly penetrate pores, tiny holes, cracks, or other sources of leakage.
The inspected parts are not soiled or contaminated. Therefore, rust protection, drying, or cleaning is not necessary.
A leak test is only possible if the test specimen is adapted using sealing elements or fixtures. HeMaTech also offers a customized solution for this—just ask us.
Operation
Test parameters are entered into a clear program schedule and require no special prior knowledge. On-screen help is available if needed. Data security is ensured by a write-protect key switch. When the device is locked, the parameters can only be read but not changed.
Higher-level control systems can communicate with the 70PRO via a standard digital I/O interface or, optionally, via Profibus or Profinet. Test data and program parameters can also be exchanged via Profibus/Profinet. The 70PRO can therefore be easily integrated into automated manufacturing processes.
Test programs can be entered directly into the test device very easily via the touchscreen or an optional connected keyboard. The devices also feature a built-in VNC server. Once connected to a network, the device can be accessed and operated via its IP address and a VNC client.
The front-panel USB port allows for easy storage of measurement data and test programs from the leak tester.
Pneumatic Section Design
- Industrial-grade, rugged differential pressure transmitter
• Overpressure protection up to 150 bar/g
• Temperature-compensated
• Easy to maintain
- Electronic pressure control for filling and test pressure
- Pneumatically controlled isolation and shut-off valve (not affected by valve heating)
- Function monitoring
- Leak simulation port for connecting an external leak calibrator or test leak
Electronic Section Design
- Windows CE-based system
- Data Backup in Persistent NAND Flash Memory
- Memory for 99 programs and 10,000 data records
- 7-inch touchscreen display
- virtual keyboard on the touchscreen
- Remote Control via VNC
- Write-protect switch to prevent unauthorized changes to the parameters,
- Plain text error messages
- Easy Programming
- Various interfaces for communication with higher-level control systems
Each data record contains information on the test program, date, time, measured value, test result, errors, and test parameters.
Larger amounts of data can be stored on an external USB storage device connected to the front panel.
External keyboard via USB
VNC Viewer
, 800 × 480 px
0.10 ... 1.0 bar(g)
0.15 ... 2.5 bar(g)
0.20 ... 6.0 bar(g)
0.50 ... 10.0 bar(g)
-0.80 ... -0.10 bar(g)
Combined pressure ranges consisting of positive and negative pressure
Other pressure ranges available upon request
Test pressure: … bar
Filling time: 0.1 ... 999.9 s
Equilibration time: 0.1 ... 999.9 s
Measurement time: 0.1 ... 999.9 s
Rework starting at: …
Leakage starting at: …
Serial defects starting at: … “Leaky” parts
, Rework
, No leak
, Malfunction
Control inputs: 24 V, 500 mA
Control outputs: 24 V, 500 mA, short-circuit-proof
Maximum 10 bar(g)
Dry, oil-free, and filtered
Test medium connection: G1/4"
Optional test connection: G1/8"
Reference volume connection: G1/8"
Venting via silencer: G1/4"
Key for the write-protect switch
USB flash drive containing the user manual and documentation